The AA8000 Atomic Absorption Spectrophotometer is a double beam with sealed and vibration free optical system, having Czerny-turner Monochromator with Holographic grating (1800 lines/mm). It’s a high performance automated instrument designed to meet the requirements of the modern laboratory. Due to its versatility and performance it can be used for a wide range of applications including: Agriculture, Soil, Environmental, Food, Metal, Mining, Petrochemical, Clinical, Pharmaceutical.
Standard Features of Fully Automatic True Double Beam – Atomic Absorption Spectrophotometer (AA8000)
Benchtop Model.
Software Features
Gas Flow Control: Fully Automatic, Software operated mass flow control for Fuel Flow. Total gas flow control system to maintain gas flow (Ratio of fuel/Oxidant) at set levels even when subjected to outside variations like nebulizer adjustments. Automatic gas flow adjustment during change over between Air / C2H2 & N2O.
Typical Specification
Wavelength Range | 185nm – 915nm |
Optics | True double Beam, Sealed & Vibration free optical system with a reflective optical compartment. |
Detector | PMT (wide range) |
Monochromator | Czerny-turner type with Holographic Grating of 1800 lines/mm and Reciprocal Linear Dispersion better than 1.6 nm/mm, focal length 300mm. |
Spectral Bandwidth | Adjustable slit from 0.1 nm. to 2 nm. (software selectable) |
Wavelength Accuracy | ±0.15nm |
Wavelength Reproducibility | < 0.05nm |
Resolution | 0.2nm ± 0.02nm |
Baseline stability | 0.005A/30min |
Sensitivity(Cu) | >0.9 Absorbance or better for 5 ppm Cu solution |
Detection Limit | Cu < 0.004 μg/ml (flame) Cd < 0.4 x 10-12 g (graphite furnace) |
Repeatability | Cu < 0.7% ( Air/Acetylene flame)
Ba < 1.0% ( Nitrous oxide/Acetylene flame) Cu < 2.0% Cd < 2.0% (Graphite Furnace) |
Background Correction | Deuterium Arc / Self reversal upto 2.5 Absorbance |
Characteristic correction | Cu < 0.02 μg/ml, Ba < 0.15 mg/ml (NO/Acetylene) Burner Heads |
Nebuliser | High Efficiency Nebuliser |
Burner Head | Titanium Alloy |
Atomization Chamber | Corrosion-resistant material |
Position Adjustment | Automatic changeover (AA8000GF) Manual (AA8000F)
Automatic Setting of Optimum Height for Flame Burner. Burner movement in horizontal & vertical direction by software. |
Safety Functions (Interlocks) | Burner Identification, Flame Sensor, Gas leak Sensor, Low Gas Pressure Sensor, Drain Trap Sensor, Power Loss Protection, Circulation Water (graphite), Over Temperature Sensor (graphite) |